T502P

Halfbridge Transducer with 7.38 mm/Vmm sensitivity and ± 5 mm measuring stroke, Pretravel adjustable. Pneumatic push. Cable 2 m, cable exit radial.

Articlenumber
1001860
Product Image
Properties
  • Total Stroke, Measuring stroke (symmetrical ±), Pretravel, Pretravel default setting, Bearing, Lifetime, Tip rotation, Temperatur range, Mounting position, Tip, Bellow, Body diameter, Plug, Cable
    Total Stroke
    10.6 mm
    Measuring stroke (symmetrical ±)
    5.0 mm
    Pretravel
    adjustable
    Pretravel default setting
    symmetrical
    Bearing
    Ball bearing, no side-play
    Lifetime
    > 10 Mio. Cycles, bellow wear part exchangeable
    Tip rotation
    1° over full stroke
    Temperatur range
    -10 to +65 °C, storage and operation
    Mounting position
    any
    Tip
    3 mm tungsten carbide ball, M2.5 fixing thread, exchangeable
    Bellow
    FPM / FKM
    Body diameter
    8h6
    Plug
    5 pin, 240°
    Cable
    PUR shielded, Length 2 m
  • Advance, Lift off, Pressure Pmax, Pressure P application, Spring rate information
    Advance
    pneumatic
    Lift off
    return spring
    Pressure Pmax
    1.5 bar
    Pressure P application
    1.2 bar
    Spring rate information
    apx. 2.0 N at 1.2 bar / apx. 1.5 N at 0.9 bar (at el. zero)
  • Repeatability, Linearity error
    Repeatability
    0.02 µm
    Linearity error
    0.8% FS ± 5000 µm range (at 20 °C ± 1°C)
  • Sensitivity, Optional sensitivity setting, Setting, Setting optional, Amplifier input load, Supply voltage, Drive frequency, Coil scheme
    Sensitivity
    7.38 mV/(Vmm)
    Optional sensitivity setting
    14.76 mV/(Vmm)
    Setting
    1:10
    Setting optional
    1:5
    Amplifier input load
    2 kOhm Input resistance, ± 0.1%
    Supply voltage
    3.0 V ±0.5%RMS
    Drive frequency
    13.0 kHz ±5%
    Coil scheme
    Halfbridge, TESA® compatible
  • Documents information
    Documents information
    All drawings and 3D models are shown in "electrical zero" position.